Measurement
45
User check menu
Entering the user check menu allows you to view the display and transmission output and to
initialize the settings.
Entering the user check menu
1.
Hold down HOLD in the measurement mode until the HOLD indicator turns ON.
2.
Use the arrow keys [
/
] and select “USr” for the user check menu.
3.
Press ENT.
4.
Use the arrow keys [
/
] to select the desired item displayed on the sub display.
5.
Press ENT.
Table 6 List of items displayed for the user check menu
Display
Item
Description
Screen to check the LCD
Screen to check the LCD
The whole LCD is illuminated, allowing you to check the display.
Screen to check the output
Screen to check relay
output 1
ON/OFF check of relay 1 (r1).
ENT: Alternates ON and OFF.
Note: When this mode becomes active, the contact output turns OFF.
Screen to check relay
output 2
ON/OFF check of relay 2 (r2).
ENT: Alternates ON and OFF.
Note: When this mode becomes active, the contact output turns OFF.
User check menu
Press HOLD or ESC
for a while.
Measurement mode
Hold mode
Press HOLD for a while.
Display of various
settings
Setting
change
Establishes
the value
Cancels
the value
h19136 | UTC 2020/03/30 3:34:51