Installation
21
Contact output
The contact capacity is a resistance load for 250 V AC, 3 A max. or 30 V DC, 3 A max.
The terminal screw for the contact output is M4.
The applicable electric wire is 0.75 mm
2
to 5.5 mm
2
(AWG 18 to AWG 10).
If noise is detected from the load, use a varistor or a noise killer.
When the HR-200 is normal (without the FAIL output), the C-NOF contact is open and the
C-NCF contact short-circuited. When the power is OFF, the C-NOF contact is short-
circuited.
Connect the contact output as illustrated below.
Transmission output
The terminal screws for the contact input and the transmission output are M3.5.
The applicable electric wire is 2 mm
2
(AWG 14) max.
For the transmission output cable, use a shielded cable.
WARNING
WARNING
NOF
CF
NCF
Not assigned
FAIL (error warning output)
NC1
C1
NO1
NC2
C2
NO2
R1 (control output) R2 (control output)
Not used for the HR-200.
Do not connect anything.
(+) (
)
(+) (
)
OUT1
OUT2
(transmission output)