Chapter 3 Basic Operation
3.3 Calibration
68
HORIBA
6.
Press the
or
key to select digit and press the
▲
or
▼
key to change value, to set the value of the
desired standard solution.
7.
Check the stable indication value, and press the
CAL key.
The HOLD indicator blinks until the reading
stabilizes. When the reading stabilizes, the HOLD
indicator lights up and the reading is held.
8.
After the calibration is completed, press the MEAS
key to return to the measurement screen.
▲
▼