Chapter 3 Basic Operation
3.2 Settings
F-74BW
27
3.2.10 Conductivity Cell Constant Setting
Set the cell constant of electrode (Setting range: 0.700
to 1.300). When using a electrode for the instrument for
the first time, set the cell constant written on the label of
the electrode.
This setting is possible only for CH2.
1.
Press the SET key on the measurement screen,
and press the
▲
or
▼
key to enter the conductivity
cell constant setting screen.
2.
Press the ENTER key to enter the conductivity cell
constant change screen.
When the
or
key is pressed, the display
changes depending on the current unit setting.
For the SI unit (S/m)
×
10
→
×
100
→
×
1000
→
×
10
For the conventional units (S/cm, mS/cm
)
×
0.1
→
×
1
→
×
10
3.
Press the ENTER key.
▲
▼
FIX