Chapter 3 Basic Operation
3.3 Calibration
52
HORIBA
4.
Press the
▲
or
▼
key on the measurement screen
to light up ACT for CH1.
5.
Press the CAL key to enter the CUST calibration
screen.
6.
Press the
or
key to select digit and press the
▲
or
▼
key to change value, to set the value of the
desired standard solution.
7.
Check the stable indication value, and press the
CAL key.
The HOLD indicator blinks until the reading
stabilizes. When the reading stabilizes, the HOLD
indicator lights up and the reading is held.
The
mark, which means that the first point
calibration is completed, is displayed on the screen.
・
To stop calibration:
Press the CAL key while the HOLD indicator
blinks.
・
To redo calibration:
Press the CAL key again after the HOLD indicator
lights up.
▲
▼