Chapter 3 Basic Operation
3.2 Settings
26
HORIBA
3.2.9 Conductivity
Temperature Coefficient Setting
Set the coefficient used for temperature compensation in
electric conductivity measurement (Input range: 0.00%
to 10.00%). The default is 2.00%.
This setting is possible only for CH2.
1.
Press the SET key on the measurement screen,
and press the
▲
or
▼
key to enter the conductivity
unit setting screen.
2.
Press the ENTER key to enter the conductivity tem-
perature coefficient change screen.
3.
Press the
or
key to change ON or OFF, and
press the ENTER key .
4.
Press the
or
key to select digit and press the
▲
or
▼
key to change value, and ENTER key.
Setting range: 0.00% to 10.00%
▲
▼
▲
▼