![horiba ENDA-5000 Instruction Manual Download Page 97](http://html1.mh-extra.com/html/horiba/enda-5000/enda-5000_instruction-manual_2162208097.webp)
9 Technical Data
85
Calibration gas
Zero gas
N
2
or ambient air
Carrier gas for
O
2
measurement
Ambient air
Span gas
Gas cylinder for measured component
Probe
Filter at sampling point
Flange: JIS 10 K, 40 AFF
Probe tube length: 1000 mm
Material: SUS-316
Element: SUS-304+quartz wool bellows of 2 µm in thickness
Electric heating: 100 VA with drip-proof case
Power voltage
100 V ±15 V AC
Power frequency
50/60 Hz switchable
Power consumption
Approximately 800 VA (Including primary filter and service plug outlet)
Heated tube 30 m (+1100 VA) and panel heater (+300 VA)
Dimensions
Mass
600 (W) mm x 1770 (H) mm x 300 (D) mm
(High-pressure gas cylinder, maximum 3 gas cylinders of 3.4 L are contained.)
Approximately 180 kg (Without cylinders)
600 (W) mm x 1770 (H) mm x 500 (D) mm
(High-pressure gas cylinder, maximum 6 gas cylinders of 3.4 L are contained.)
Approximately 200 kg (Without cylinders)
Materials exposed to gas
SUS-316, SUS-304, Tefon, polypropylene, polyethylene, fluoro rubber, chloroethene, PVDF,
and glass
Cabinet
Standalone type for outdoor installation
Plate thickness: 2.3 mm for steel plates of main unit, door. and top plate,
3.2 mm for channel base
Door: Front
Connections: Right side
Color
Munsell 5Y7/1 semi-gloss for inner and outer surfaces
*1: The analyzer against N
2
O interference for CO analyzer applies the standard range of 200 ppm or more
(no optional range).
*2: No carrier gas cylinder is necessary.
*3: For the temperature range of -15ºC to 40ºC (cold district) and for the specification of -5ºC to 50ºC, we will separately
discuss the design.
*4: When the coexisting gas contains NH
3
, NH
3
scrubber is prepared for an optional part.
When CH
4
coexists in the sample gas for SO
2
analyzer, the SO
2
analyzer of CH
4
interference compensation type is
prepared.
When N
2
O coexists in the sample gas for CO analyzer, the CO analyzer of N
2
O interference compensation type is
prepared.