10 Optional Functions
99
10.1.3 Purge start
Q
Automatic start
Select [INTERNAL] in the PURGE MODE screen (page 95 Fig. 69).
The purge automatically starts according to the purge start time, purge interval, purge in-blow
time, and purge delay time set at " 10.1.2 Purge setting " (page 93).
Note
During the purge, the concentration output is held.
Q
Manual start
1.
Press the [MENU] key in the MEAS. screen.
Fig. 74 MENU 1/5 (OPERATION menu) screen
2.
Press the [PURGE START] button.
The purge starts according to the purge in blow time and purge delay time set at "
10.1.2 Purge setting " (page 93).
Note
When the purge starts, the [PURGE START] button changes into the [PURGE STOP] button.
Pressing the [PURGE STOP] button interrupts the purge.
During the purge interruption, the [PURGE STOP] button changes into the [PURGE START] button.
Q
Start from external contact input
Select [EXTERNAL] in the PURGE MODE screen (page 95 Fig. 69).
The purge starts according to the purge start time, purge interval, purge in-blow time, and
purge delay time set at " 10.1.2 Purge setting " (page 93) via external contact input.
Note
The purge by manual start via the [PURGE START] button is available except for during the AIC.
When the purge command is input by the internal clock or via the external contact input during the
purge of manual start mode, the signal is not received but the purge by the manual purge continues.
* Functions of “PURGE START,” “ROLLING AVERAGE RESET”
are optional.