10 Optional Functions
94
3.
Press the button corresponding to the item to be set. The setting screen appears.
For the detailed information on each setting screen, refer to page 95 to page 98.
4.
Change the setting in each setting screen, and press the [SET] key.
The setting is changed and the screen returns to the PURGE SET screen.
Tip
To cancel the change, press the [CANCEL] key.
The change is canceled and the screen returns to the PURGE SET screen.
5.
Press the [EXIT] key in the PURGE SET screen.
The screen returns to the MENU 4/5 (OPTION menu) screen.
6.
Press the [EXIT] key in the MENU 4/5 (OPTION menu) screen.
The screen returns to the MEAS. screen.
Note
Trying to change the purge setting during the purge is invalid. The screen only shows “IN PURGE.”
Setting item
Description
MODE
Sets the purge start mode.
Pressing the button at the right of [MODE] displays the PURGE MODE screen (page 95 Fig. 69).
START
Sets the start time of the next purge sequence.
Pressing the button at the right of [START] displays the PURGE START screen (page 96 Fig. 70).
INTERVAL
Sets the time span (INTERVAL) for the purge start.
Pressing the button at the right of [INTERVAL] displays the PURGE INTERVAL screen (page 97 Fig.
71).
IN BLOW TIME
Sets the purge time for the primary filter.
Pressing the button at the right of [IN BLOW TIME] displays the PURGE IN BLOW TIME screen
(page 97 Fig. 72).
DELAY TIME
Sets the time in which the sample tube is replaced from the purge gas to the sample gas.
Pressing the button at the right of [DELAY TIME] displays the PURGE DELAY TIME screen (page 98
Fig. 73).