10 Optional Functions
101
2.
Press the [O
2
CORRECTION SET] button.
The O
2
CORRECTION SET screen appears.
Fig. 76 O
2
CORRECTION SET screen
3.
Press the button corresponding to the item to be set. The setting screen appears.
For the detailed information on each setting screen, refer to page 102 to page 105.
4.
Change the setting in each setting screen, and press the [SET] key.
The setting is changed and the screen returns to the O
2
CORRECTION SET screen.
Tip
To cancel the change, press the [CANCEL] key. The change is canceled and the screen returns to the
O
2
CORRECTION SET screen.
5.
Press the [EXIT] key in the O
2
CORRECTION SET screen.
The screen returns to the MENU 4/5 (OPTION menu) screen.
6.
Press the [EXIT] key in the MENU 4/5 (OPTION menu) screen.
The screen returns to the MEAS. screen.
Setting item
Description
MODE
Two types of O
2
set values: O
2
SET VALUE1 and O
2
SET VALUE2 can be selected.
The switchover between two is available via the external contact input.
Pressing the button at the right of [MODE] displays the O
2
CORRECTION MODE screen (page
102 Fig. 77).
O
2
SET VALUE 1
Sets the O
2
SET VALUE 1.
Pressing the button at the right of [O
2
SET VALUE 1] displays the O
2
SET VALUE 1 screen (page
103 Fig. 78).
O
2
SET VALUE 2
Sets the O
2
SET VALUE 2.
Pressing the button at the right of [O
2
SET VALUE 2] displays the O
2
SET VALUE 2 screen (page
104 Fig. 79).
O
2
LIMIT VALUE
The O
2
correction cannot obtain the operation accuracy when the O
2
measurement value is close
to 21 vol%. When the value exceeds the set limit value, the O
2
limit value is operated as the O
2
measurement value.
Pressing the button at the right of [O
2
LIMIT VALUE] displays the O
2
LIMIT VALUE screen (page
105 Fig. 80).