8 Troubleshooting
78
Note
z
The alarm/caution messages marked by “*” are not output as analyzer alarm/analyzer caution
contact output when they occur within 1 hour from power ON.
They are not recorded on the alarm history. Only the presence history of alarms/cautions can be
confirmed from the screen. Refer to " 6.7 Alarm Check " (page 54).
z
Analyzer Alarm (contact output):
Indicates that the analyzer is in the state disabling the measurement or the state hard to maintain
the measurement accuracy. Handle it as soon as possible.
z
Analyzer Caution (contact output):
Indicates that the measurement accuracy is maintained but starting to be hard to be maintained.
Confirm the situation and prepare for the countermeasures.
Span
Calibration
Caution
Indicates that the
coefficient revised at the
span calibration starts
reaching the outside of
specified range.
Confirm the gas type
of calibration gas
cylinder.
Confirm that there is
enough residual
pressure of
calibration gas.
By the flowmeter of
analyzer front panel,
confirm that there is
the specified flow.
If the wrong gas type is
selected, perform the
calibration again with the
correct gas type.
If the residual pressure of
calibration gas is within 1
MPa, replace it into a new
cylinder and perform the
calibration again.
Adjust the flow so that it is
within the specified level,
and perform the calibration
again.
If the situation is not
improved, contact HORIBA.
page 19
page 30
Analyzer caution
I2C
COMM ERR
IDn (n: 0 to 7)
Indicates that the com-
munication system
inside the instrument
has abnormalities.
-
Contact HORIBA.
-
-
HI.HI
(optional)
Indicates that the
measurement
concentration value is
over the HI HI limit
alarm set value.
-
-
-
Concentration
HI HI limit
HI
(optional)
Indicates that the
measurement
concentration value is
over the HI limit alarm
set value.
-
-
-
Concentration
upper limit
LO
(optional)
Indicates that the
measurement
concentration value is
below the LO limit alarm
set value.
-
-
-
Concentration
lower limit
CO Peak
Count
(optional)
Indicates that the peak
count is over the set
value for CO analyzer.
-
-
-
CO peak count
Item
Description
Check item
Countermeasure
Reference
Contact output