10 Optional Functions
112
3.
Press the button corresponding to the item to be set. The setting screen appears.
For the detailed information on each setting screen, refer to page 112 to page 114.
4.
Change the setting in each setting screen, and press the [SET] key.
The setting is changed and the screen returns to the MEASURE VALUE ALARM
screen.
Tip
To cancel the change, press the [CANCEL] key. The change is canceled and screen returns to the
MEASURE VALUE ALARM screen.
5.
Press the [EXIT] key in the MEASURE VALUE ALARM screen.
The screen returns to the MENU 4/5 (OPTION menu) screen.
6.
Press the [EXIT] key in the MENU 4/5 (OPTION menu) screen.
The screen returns to the MEAS. screen.
Note
z
The following state stops the alarm judgment and releases an alarm in occurrence.
・
5 minutes from power ON
・
In maintenance
z
The following state maintains the condition just before the alarm occurrence.
・
In AIC (In introducing the calibration gas)
・
In purge
Q
Component
Pressing the button showing the component to be set in the MEASURE VALUE ALARM
screen displays the COMPONENT screen.
Fig. 89 COMPONENT screen (MEAS. mode)
Setting item
Description
COMPONENT
Sets the component for which the alarm is output.
Pressing the button at the lower of [COMPONENT] displays the COMPONENT
screen (page 112 Fig. 89).
TYPE
Sets the HIGH, H.HIGH, and LOW alarm.
Pressing the button at the lower of [TYPE] displays the ALARM TYPE screen
(page 113 Fig. 91).
VALUE
Sets the concentration regarded as alarm.
Pressing the button at the lower of [VALUE] displays the ALARM VALUE screen
(page 114 Fig. 92).