8 Troubleshooting
76
Electric
Cooler
Temperature
Alarm *
Indicates that the
electric cooler has
abnormalities such as
cooling unit
deterioration or power
supply breakage.
Check that the
power lamp of
electric cooler is lit
and that the cooling
fan operates
correctly.
If the power lamp is
extinguished or the cooling
fan stops, contact HORIBA.
When the cooling fan or
radiation fin is dirty, apply
cleaning.
If the alarm is not cleared,
contact HORIBA.
-
Analyzer alarm
Electric
Cooler
Temperature
Alarm *
Indicates that the
electric cooler has
abnormalities such as
cooling unit
deterioration or power
supply breakage.
Check that the
power lamp of
electric cooler is lit
and that the cooling
fan operates
correctly.
If the power lamp is
extinguished or the cooling
fan stops, contact HORIBA.
When the cooling fan or
radiation fin is dirty, apply
cleaning.
If the alarm is not cleared,
contact HORIBA.
-
Analyzer caution
Temperature
in Cabinet
Alarm
(optional) *
Indicates that the
temperature inside
cabinet is out of the
specified range due to
the temperature
fluctuation outside.
Confirm that the
temperature at
installation part is
within the range of -
5ºC
to 50ºC.
If the specification is
with the panel
heater, confirm that
the panel heater CP
is ON.
When the temperature is
high:
Consider counteractions
such as a shield
installation.
When the temperature is
low:
Contact HORIBA.
-
Analyzer alarm
Caution for
Temperature
inside
Cabinet
(optional) *
Indicates that the
temperature inside the
cabinet is out of the
specified range.
Confirm that the
temperature at
installation part is
within the range of
-5ºC
to 50ºC.
If the specification is
with the panel
heater, confirm that
the panel heater CP
is ON.
When the temperature is
high:
Consider counteractions
such as a shield
installation.
When the temperature is
low:
Contact HORIBA.
-
Analyzer caution
Atmospheric
Sensor Alarm
Indicates that the
abnormalities have
occurred such as
atmospheric sensor
deterioration.
-
Contact HORIBA.
-
Analyzer alarm
Gas Residual
Pressure
Decreased
(optional)
Indicates that the
residual pressure of
calibration gas cylinder
has been decreased.
Confirm the residual
pressure of
calibration gas
cylinder.
If the residual pressure is
within 1 MPa, replace it into
a new cylinder.
page 19
Analyzer caution
Item
Description
Check item
Countermeasure
Reference
Contact output