4 CALIBRATION
20
INTERVAL
Set the time interval, which applies if the AIC sequence is started periodically.
Pressing the displayed INTERVAL setting will display the INTERVAL screen.
Fig. 25 INTERVAL screen
Press the value to be changed. The value will be highlighted, allowing you to change it.
Using the [
] and [
] buttons, change the value.
Note
INTERVAL should be set to the AIC sequence time plus 10 minutes or longer.
If the [SET] key is pressed with a shorter interval entered, the period equivalent to the AIC sequence
time plus 10 minutes will be set automatically.
Item
Settable range
Day
0 to 999
Hour
00 to 23
Minute
00 to 59