6 FUNCTIONALITIES
62
6.6.2
Touch panel adjustment
Press the [TOUCH PANEL ADJUSTMENT] button on the MENU/SYSTEM screen.
The TOUCH PANEL ADJUSTMENT screen will be displayed.
If the displayed buttons and keys deviate from their respective touch positions, you can adjust
the touch panel by performing the following steps.
Note
Such deviations on the touch panel can be adjusted within a range of 36 dots.
If the positions deviate exceeding this limit, the following steps cannot be performed. Contact us.
Adjustment procedure:
1.
Press the [TOUCH PANEL ADJUSTMENT] button on the MENU/SYSTEM screen.
The TOUCH PANEL ADJUSTMENT screen (1) will be displayed.
2.
Press the center of the area framed by 4 small triangles.
Fig. 76 TOUCH PANEL ADJUSTMENT screen (1)
3.
The TOUCH PANEL ADJUSTMENT screen (2) will be displayed. Press the center of
the area framed by 4 small triangles. The positions on the touch panel will be
corrected and the MENU/SYSTEM screen will be displayed again.
Fig. 77 TOUCH PANEL ADJUSTMENT screen (2)