4 CALIBRATION
26
4.4
Manual Calibration
After preparing for the calibration (see “4.2 Preparation for Calibration” (page 14)), perform the
zero calibration and the span calibration in this order.
4.4.1
Operational flow
The operational flow for manual calibration is described below:
Note
Control the zero gas and span gas feed pressures at atmospheric pressure plus (0 kPa to 0.5 kPa).
When using the calibration gas line:
When using the measured gas line
Reference page
Gas line connection check
↓
|
Span value setting
page 14
↓
↓
Measurement line switching
[MEAS.] -> [ZERO]
Measurement line check
[MEAS.]
page 27
|
↓
Zero gas introduction
from the calibration gas line
Zero gas introduction
from the sample inlet
↓
|
Waiting for readout stabilization
↓
↓
Zero calibration
↓
|
Measurement line switching
[ZERO] -> [SPAN]
|
page 28
↓
↓
Span gas introduction
from the calibration gas line
Span gas introduction
from the sample inlet
↓
↓
Waiting for readout stabilization
↓
↓
Span calibration
↓
↓
Line switching
[SPAN] -> [MEAS.]
|
↓
Sample gas introduction
from the sample inlet