4 CALIBRATION
19
LIMIT (START-END)
Set the range of time available for starting the AIC sequence.
Pressing the displayed LIMIT (START-END) setting will display the LIMIT (START-END)
screen.
Fig. 24 LIMIT (START-END) screen
Press the value to be changed. The value will be highlighted, allowing you to change it.
Using the [
] and [
] buttons, change the value.
Note
When you do not use the LIMIT (START-END) function, select the default value (00:00 to 00:00).
If the START and END values of the range are the same, the LIMIT (START-END) function is
invalid.
Item
Settable range
Start: Hour
00 to 23
Start: Minute 00 to 59
End: Hour
00 to 23
End: Minute 00 to 59