The EPA designation applies to G, -1, G-1, and later APDA-371 PM10 Beta Attenuation Monitors, when used in
conjunction with the following requirements. Users are advised that configurations that deviate from this specific
description may not meet the applicable requirements of 40 CFR Parts 50 and 53.
•
The APDA-371 is operated to obtain a daily average of the hourly measurements, with a filter change frequency
of one hour.
•
The inlet must be equipped with the standard BX-802 EPA PM
10
inlet head.
•
The unit must be used with standard glass fiber filter tape.
•
The unit may be operated with or without any of the following options: BX-823 inlet tube extension, BX-825
heater kit, BX-826 230V heater kit, BX-828 roof tripod, BX-902 exterior enclosure, BX-903 exterior enclosure
with temperature control, BX-961 mass flow controller, BX-967 internal calibration device.
•
The SAMPLE TIME parameter must be set for 50 minutes.