This file will contain text of all of the data stored by the APDA-371, since the last time the data was downloaded. Useful
to avoid duplicate data in your database. A flag is set in the APDA indicating where the last download stopped.
File 4:
Display System Configuration
This file will contain a list of most of the APDA-371 settings and calibration values as shown below. This is useful for
verifying the setup on a remote APDA, and to send to the factory if service is required. The setting report has been
updated and reformatted. Following is an example of the new settings report. Older firmware will show a slightly
different report.
APDA 1020 Settings Report
06/07/2007 14:19:45
Station ID, 1
Firmware,
3236-02 3.2.5
K,
01.000
BKGD,
00.000
usw,
00.301
ABS,
00.805
Range,
1.000
Offset,
-0.015
Clamp,
-0.015
Conc Units,
mg/m3
Conc Type,
ACTUAL
Cv,
01.000
Qo,
00.000
Fow Type,
ACTUAL
Flow Setpt,
0016.7
Std Temp,
25
Temp Mult,
1.0000
Pres Mult,
1.0000
Flow Mult,
1.0000
High Flow Alarm,
20
Low Flow Alarm,
10
Heat Mode,
AUTO
Heat OFF,
20
RH Ctrl,
YES
RH SetPt,
35
RH Log,
YES
DT Ctrl,
NO
DT SetPt,
99
DT Log,
NO
APDA Sample,
42
MET Sample,
60
Cycle Mode,
STANDARD
Fault Polarity,
NORM
Reset Polarity,
NORM
Maintenance,
OFF
EUMILRNFPDCT
000000000000
AP,
000150
Baud Rate,
9600
Printer Report,
2
e3,
00.000
e4,
15.000