INITIAL SETUP OF YOUR APDA-371
This section describes the process for setting up and configuring your APDA-371, as well as the basic steps required to
put the unit into operation. Some of the topics in this section will direct you to other sections of this manual for more
detailed information. It is assumed that the unit is already installed and sited as described in Section 2. In some cases it
is useful to first set up the APDA-371 unit on a test bench before deployment or installation in order to explore the
functions of the unit and perform setups. The following steps for starting up your unit are described in this section:
1.
Power on and warm up the unit.
2.
Familiarize yourself with the user interface.
3.
Load a roll of filter tape.
4.
Perform a Self-Test.
5.
Set the real-time clock, and review your SETUP parameters.
6.
Perform a leak check and a flow check.
7.
Return to the top-level menu and wait for automatic start at the top of the hour.
8.
View the OPERATE menus during the cycle.
3.1
Power On
The APDA-371 has a power switch located on the back of the unit directly above the power cord. Verify that the unit is
plugged in to the correct AC voltage, and that any electrical accessories are correctly wired before turn the unit on.
(Section 2.6) When power is switched on the main menu screen should appear after a few seconds as shown below.
The unit will probably flash an error indicating that there is no filter tape installed. Note: Units running revision 3.1 or
earlier firmware will display a slightly different main menu screen.
The Main
Menu Screen
3.2
Warm-up
The APDA-371 must warm up for at least one hour before an operation cycle is started. This is because the beta
detector contains a vacuum tube which must stabilize every time the unit is powered up. This also allows the electronics
to stabilize for optimal operation. This applies any time the unit is powered up after being off for more than a moment.
Instrument setups and filter tape installation can be performed during this warm up time. Most agencies choose to
discard the first few hours of concentration data after the APDA is powered up.
3.3
Using the Keypad and Display
When the APDA-371 is powered up it will display the main (top level) menu on the LCD display. This menu is the
starting point for all functions of the APDA-371 user interface.
3/28/2007 BAM 1020 16:08:29
LAST CONCENTRATION: 0.028 mg/m3
CURRENT FLOW: 16.7 LPM
STATUS: ON
FIRMWARE: 3236-02 3.2.4
SETUP OPERATE TEST TAPE
F1
F2
F3
F4
F5
F6
3/28/2007 BAM 1020 16:08:29
LAST CONCENTRATION: 0.028 mg/m3
CURRENT FLOW: 16.7 LPM
STATUS: FILTER TAPE ERROR!
FIRMWARE: 3236-02 3.2.4
SETUP OPERATE TEST TAPE