9. If the lowest (initial) value of the measuring range is at zero (for example in “0 to
100”) then do not change this value from its default of 0.
10. Once the desired lower value is showing on the display, hold the magnet over the (
)
switch for one second and remove. The display will then show ‘HIgh 100’ (the ‘100’
part is flashing). ‘HIgh’ means the highest value in any given measuring range.
11. Now with the magnet, use the (▲) or (▼) switches to select the desired upper value
of the measuring range.
12. Once the desired upper value is showing on the display, hold the magnet over the (
)
switch for one second and remove.
13. The display will then show ‘CAL YES’ (the ‘YES’ part is flashing).
14. Now hold the magnet over (
) for one second and remove to begin calibration of the
sensor.
Go to page 32, step 5.