background image

GENESIS

TM

 DM MONITOR

 

Programming/Troubleshooting Guide 

 

P4520EN.04 07/14/2010

   

 

 

 

 

22

 

VFD Display 

Meaning 

Possible Cause 

Resolution/Action 

324 PAGER_MSG_FAIL 

Message 

failed 

 

If problem persists, call 
Honeywell HomMed. 

350 GPRS 

Requires 

Reset 

 

Modem responded with an error 
the monitor doesn't handle, too 

early to be more specific with 
causes here 

If problem persists, call 
Honeywell HomMed 

351 

Call Customer Service 

Improper Hardware Setup 

SIM card is missing or not 
enabled 

If problem persists, call 
Honeywell HomMed 

352 

GPRS Service Unavailable 

Signal is too weak for the 
modem to work 

No service in the area 

If problem persists, call 
Honeywell HomMed. 

353 

GPRS Network Fail 

Access Point Name is not 
configured correctly on the 

modem 

No service in the area 

If problem persists, call 
Honeywell HomMed. 

410 GLUCO_TIMEOUT 

User timeout – glucose 
meter 

Patient noncompliance 

Retrain the patient.  

411 

GLUCO_COMM 

Glucose meter disconnected 

Glucose meter cable not 

connected 

Loose Connection 

Connect the cable.  

 

Make sure all connections are 
secure. 

412 

GLUCO_OFF 

Glucose meter not enabled  

 

In Options menu, set appropriate 

glucose meter to On.  

413 WRONG_GLUCO 

ID Card and monitor setup 
gluco-type mismatch 

 

Program the card or the monitor 
for the correct glucose meter.  

420 

SPIRO_TIMEOUT 

User timeout - spirometer 

Patient noncompliance 

Retrain the patient.  

421 SPIRO_COMM 

Spirometer 

disconnected 

Spirometer cable not connected 

Loose connection 

Connect the cable. 

 

Make sure all connections are 
secure. 

422 SPIRO_OFF 

Spirometer not enabled (no 

assigned COM port) 

 

In Options menu, set appropriate 

spirometer to On.  

430 

PT_TIMEOUT 

User timeout – PT/INR 

Patient noncompliance 

Retrain the patient.  

431 PT_COMM 

PT/INR 

disconnected 

PT/INR cable not connected 

Loose connection 

Connect the cable. 
 

Make sure all connections are 
secure. 

432 PT_OFF 

PT/INR not enabled (no 
assigned COM port) 

 

In Options menu, set appropriate 
PT/INR to On. 

440 

ERROR_ECG_TIMEOUT 

User timeout – ECG  

Patient noncompliance 

Retrain patient. 

441 ERROR_ECG_COMM 

ECG 

disconnected 

ECG cable not connected 

Loose connection 

Connect cable.  

Make sure all connections are 
secure. 

442 ERROR_ECG_OFF 

ECG not assigned a COM 

port 

 

In Options menu, set ECG to On. 

450 

PEAKFLOW_TIMEOUT 

User timeout – peak flow 

Patient noncompliance 

Retrain the patient. 

Summary of Contents for HomMed Genesis

Page 1: ...oneywell HomMed Customer Service Toll free 888 353 5404 Phone 262 252 5840 Fax 262 252 6105 Honeywell HomMed Corporate Offices Honeywell HomMed 3400 Intertech Drive Suite 200 Brookfield WI 53045 Phone 262 783 5440 Fax 262 252 6105 Copyright 2010 Honeywell HomMed All rights reserved ...

Page 2: ...mer Service will make every attempt to resolve the issue over the phone Special Conventions The following table lists the conventions used in this guide This convention Identifies Function font Keys that you press on the keyboard sign between key names Keys that you press simultaneously START YES Courier font Text that displays in the monitor display Italics Menu names publication titles or refere...

Page 3: ... 4 Gray MANUAL key Gray MANUAL key enters Manual Mode Exits a menu in Setup Mode 5 Weight Display Displays weight from attached scale in pounds or kilograms as configured 6 Blue RETEST key Blue RETEST key starts an unscheduled monitoring session Selects a menu option in Setup Mode 7 Black NO key Black NO key records a no answer DOWN scroll key in Setup Mode 8 White YES key White YES key records a ...

Page 4: ...duling options OFF Every day Su Mo Tu We Th Fr Sa Any combination of days _ _ Mo _ _ We _ _ Fr _ _ Every time Su Only once per day S If you choose one time per day day marked with an asterisk M the monitor collects the scheduled data during the first monitoring session whether initiated by a report time or by pressing RETEST Once the scheduled data is collected the monitor will not collect it agai...

Page 5: ...GENESISTM DM MONITOR Programming Troubleshooting Guide P4520EN 04 07 14 2010 5 Programming menu chart The programming menu chart below illustrates the location of each of the monitor menus ...

Page 6: ...ing your inhalers more than usual Increased need for inhalers signals n a change in condition Use inhalers AS PRESCRIBED If difficult to breathe chest tight or in pain or sudden cough that continues Contact your CARE PROVIDER 04 Did you need extra pillows to sleep comfortably last night Using extra pillows to sleep may n signal change Use oxygen inhalers AS PRESCRIBED Contact your CARE PROVIDER EX...

Page 7: ...aving any headaches 15 Have your ankles been swollen more than usual Ankle swelling caused by What you eat medicines you take Elevate legs Limit salt IF ORDERED Call CARE PROVIDER Feet or legs more swollen 16 Are you having any difficulty standing or walking Difficulty standing or walking At risk for falls Learn cause what to do Discuss with Care Provider 17 Have you been having any new numbness o...

Page 8: ...g difficulty taking care of your wound 33 Have your hands or face been more swollen in the last day 34 Has there been any protein in your urine in the last day 35 Do you have severe heartburn today 36 Have you noticed a decrease in your baby s movement today 37 Have you had any blurred vision today 38 Have there been any changes in the medication you are taking 39 Are you having difficulty followi...

Page 9: ...ing your activities may signal a change in your medical condition If you are unable to resume normal activities have difficulty breathing or suddenly develop a new cough contact your health care provider 54 Are you having difficulty taking any of your medications same as Question Number 8 Difficulty taking medications taking medications at irregular times or not taking medications consistently may...

Page 10: ...uld be discussed with your health care provider 60 Have you developed a cough same as Question Number 28 A newly developed cough may signal a change in your medical condition If your cough is unrelieved with prescribed medication or if you develop difficulty breathing contact your health care provider 61 Have you had a fall in the last day same as Question Number 31 Falls can be a major cause of i...

Page 11: ...a meeting this week 112 Remember your grandchildren are coming over this week have fun 113 Please remember to change your dressing today 114 Please remember to keep feet and legs elevated while sitting 115 Please remember to eat foods low in Sodium 116 Please remember to wear your oxygen as prescribed 117 Please remember to use your nebulizer as prescribed 118 Please remember to do your exercises ...

Page 12: ...early every day during your sleep 7 Has anyone noticed that you quit breathing during sleep 3 4 times a week 8 Do you feel tired or fatigued today after your sleep 9 Do you feel tired fatigued or not up to par during your wake time nearly every day 10 Do you feel tired fatigued or not up to par during your wake time 3 4 times a week 11 Do you nod off or fall asleep while driving a car nearly every...

Page 13: ... the wall jack 2 Insert duplex jack 3 Plug caller ID box phone cord into duplex jack 4 Plug the monitor phone cord into duplex jack Patient already has a duplex jack in the jack outlet Replace with triplex jack 1 Remove duplex jack and insert triplex jack 2 Plug patient s equipment into two of the jack outlets and use third outlet for the monitor Have two short phone cords but need a longer cord U...

Page 14: ...urface 4 Reattach the scale cable to the scale 5 Reboot the monitor Scale does not zero properly monitor displays dashes instead of 0 0 If scale still does not zero it must be recalibrated Call Honeywell HomMed Customer Service Patient decompensates easily after obtaining weight The oximeter obtains an SpO2 reading in either 8 beats or 8 seconds The reading is taken again if it is out of range on ...

Page 15: ...rrectly positioned Poor patient perfusion Fingernail polish or false nails are interfering with sensor light Defective sensor or cable 1 Plug oximeter into plug receptacle and make sure it firmly latches into place 2 Instruct the patient on proper placement 3 Reposition sensor on patient 4 Have patient remove finger nail polish or false nails 5 Replace finger sensor Heart rate is very low or does ...

Page 16: ...gramming the monitor Verify that the cable is firmly connected to the monitor as well as the device No data transferred from device to monitor Monitor does not have a port programmed for the device Assign a com port to the device using the Option menu Communication errors between device and monitor Noise Cable problems Place the monitor so that it is not on near an electrical device such as a TV o...

Page 17: ... voice communications over a network such as the Internet or other packet switched networks Other terms frequently encountered and synonymous with VOIP include but are not limited to IP telephony Internet telephony voice over broadband VoBB broadband telephony and broadband phone Internet telephony refers to communications services voice facsimile and or voice messaging applications that are trans...

Page 18: ... numbers Verify that Transmit Mode is programmed correctly for your area If you do not know the correct mode contact Customer Service Monitor collected the vitals but did not transmit Wall jack does not work Faulty duplex jack Monitor serial number does not match unit label Monitor must dial a prefix for an outside line Facility has digital line Test wall jack with a phone Use a new duplex jack Go...

Page 19: ... SCALE_NOT_ CONFIGURED Scale not configured or configuration not sensed Scale not configured through the Setup Wizard but is turned ON in the Menu Settings Re boot re set the system first power OFF both the Scale and the DM Monitor next power ON the DM monitor first then the Scale 210 NIBP_TIMEOUT User timeout NIBP Patient noncompliance Retrain for patient 211 NIBP_COMM Lost communication with NIB...

Page 20: ...rror Call Honeywell HomMed 244 TEMP_RECOVERABLE Temperature error Probe not positioned properly Retrain the patient 245 TEMP_NO_PROBE Temperature probe disconnected Connect the probe Verify all connections secure 246 TEMP_HI_AMBIENT Temperature probe too warm Operating range 84 to 108 F 28 9 to 42 2 C 247 TEMP_LO_AMBIENT Temperature probe too cold Operating range 84 to 108 F 28 9 to 42 2 C 250 IDC...

Page 21: ...ove the block to allow the monitor to send data by modem 305 MODEM_BAUD_RATE Baud rate lock failed If using remote gateway transmission method ensure the number is correct and the remote gateway host is properly configured If problem persists call Honeywell HomMed 311 Modem Logon Skytel logon not accepted If problem persists call Honeywell HomMed 312 Modem Disconnect Skytel forced disconnect If pr...

Page 22: ...e 412 GLUCO_OFF Glucose meter not enabled In Options menu set appropriate glucose meter to On 413 WRONG_GLUCO ID Card and monitor setup gluco type mismatch Program the card or the monitor for the correct glucose meter 420 SPIRO_TIMEOUT User timeout spirometer Patient noncompliance Retrain the patient 421 SPIRO_COMM Spirometer disconnected Spirometer cable not connected Loose connection Connect the...

Page 23: ...ists call Honeywell HomMed 603 Call Customer Service User name or password was incorrect The monitor has not been activated with the ISP by Honeywell HomMed staff The ISP configuration has been modified and is no longer correct Internal and external serial number do not match If problem persists call Honeywell HomMed 604 Call Customer Service Failed communications with modem If problem persists ca...

Page 24: ... properly DSL is not configured properly The ISP is not properly accepting calls Verify the ISP1 Phone and ISP2 Phone number are configured properly in the ISP Settings menu If problem persists call Honeywell HomMed 612 ISP Login Failed Failed to log into the ISP The phone number is not configured properly The ISP is not properly accepting calls Verify the ISP1 Phone and ISP2 Phone number are conf...

Page 25: ...dia card 915 DISK_ERROR_CHANGED_CARD SmartMedia failure card swapped Reboot the unit if problem persists call Honeywell HomMed 916 DISK_ERROR_CARD_FAILURE SmartMedia failure card failure Replace SmartMedia card if problem persists call Honeywell HomMed 930 PHOTOCARD TIMEOUT No response card removed or defective Card removed before monitor could download images Defective card Retrain the patient Us...

Page 26: ...r Resources Genesis DM User Guide Honeywell HomMed LLC 2009 Genesis DM Monitor Quick Setup Guide Honeywell HomMed 2007 LifeStream User Guide Honeywell HomMed 2009 For more information about Honeywell HomMed authorized peripheral devices refer to the appropriate manufacturer s material ...

Reviews: