GENESIS
TM
DM MONITOR
Programming/Troubleshooting Guide
P4520EN.04 07/14/2010
21
#
VFD Display
Meaning
Possible Cause
Resolution/Action
301
MODEM_HARDWARE
Modem failure
Call Honeywell HomMed.
302 MODEM_TIMEOUT
No answer after successful
dial
Modem dialing wrong number
Program correct phone number.
303
No communications found
check the modem connection
or plug in the GPRS antenna
No dial tone
Modem not connected
Line is in use
Connect monitor modem to
analog phone line.
Instruct patient not to use phone
during monitoring session.
304 MODEM_CONNECT
Modem failed to connect (no
modem on other end)
Modem dialing wrong number
Phone has 800/900 block
Program correct phone number.
(Refer to Programming section in
your Sentry or Genesis
Monitoring System Clinician
Guide.)
Check for block by dialing 1-800-
679-2778. If you do not hear a
sound similar to a fax, the patient
has an 800/900 block on the
phone line. Either set the monitor
to Page Only, or explain that the
patient will need to remove the
block to allow the monitor to
send data by modem.
305
MODEM_BAUD_RATE
Baud rate lock failed
If using remote gateway
transmission method, ensure the
number is correct and the
remote gateway host is properly
configured.
If problem persists, call
Honeywell HomMed.
311
Modem Logon
Skytel logon not accepted
If problem persists, call
Honeywell HomMed.
312
Modem Disconnect
Skytel forced disconnect
If problem persists, call
Honeywell HomMed
313
Modem Go Ahead
Skytel go ahead not received
If problem persists, call
Honeywell HomMed.
314
Modem Acknowledge
Skytel packet reception not
acknowledged
If problem persists, call
Honeywell HomMed.
315
Modem Rule Violation
Skytel system rule violation
If problem persists, call
Honeywell HomMed.
316
Modem Logoff
Skytel premature logoff
initiated
If problem persists, call
Honeywell HomMed.
320
PAGER_COMM
Pager self test failure
Call Honeywell HomMed.
321
PAGER_OUT_OF_RANGE
Pager is out of range
No pager coverage in the area
Set Transmit Mode to Modem.
322
PAGER_ERR_HARDWARE
Pager failure
Call Honeywell HomMed.
323 PAGER_ERR_TIMEOUT
TX took too long (>15
minutes)
Set Transmit Mode to Modem,
and retake vitals.
If problem persists, call
Honeywell HomMed.