3Dimensions System User Guide
Chapter 6: Images
Page 80
MAN-05085-002 Revision 002
DRAFT
Preview Copy-Generated May 30, 2018
3.
To begin the contrast dose and start the timer, select the
Start
(arrow) button.
The timer begins in the
Waiting Period
, which appears with a yellow background.
Note
You can adjust the default length of time for the
Waiting Period
and
Optimal Imaging
Period
phases of the timer. Refer to
How to Set the Contrast Defaults
on page 119.
Figure 51: I-View 2D Contrast Screen, Waiting Period
Note
The timer function does not allow you to stop the timer, only to start and reset it. The
timer stops only when you exit the patient procedures.
After the
Waiting Period
, the timer begins the
Optimal Imaging Period
, which
appears with a green background.