1.2 Installation Conditions
1 - 2
1.2 INSTALLATION CONDITIONS
Check that the following conditions are satisfied for the installation
of the spectrophotometer.
1.2.1 Power Requirements
Power supply : AC 100, 115, 220, 230 or 240 V
voltage
Fluctuation should be within
10% of rated
voltage.
Frequency
: 50 or 60 Hz
Fluctuation should be within
0.5Hz of rated
frequency.
Power source : 300 VA or more
capacity
It is recommended to provide a capacity of 500
VA or more in consideration of joint use of
associated accessory devices.
Grounding line : Grounding resistance should be 100
or less.
1.2.2 Installation Place
Floor area
: Width of 500 mm and depth of 605 mm or
more. Give space at least 200 mm on
both sides of the main unit.
Work table strength : Flat-top table capable of withstanding
weight of 29 kg or more.
1.3 INSTALLATION ENVIRONMENT
1.3.1 Operating Temperature Range
5 to 35
C
In order to perform measurement under the most stable condition,
we recommend that the instrument be used in an air-conditioned
room at 20 to 25
C.
1.3.2 Operating Humidity Range
45 to 85%
1.3.3 Storage Temperature Range
-20 to +60
C