1.7 Check of Power Voltage and Fuse
1 - 6
1
.7 CHECK OF POWER VOLTAGE AND FUSE
WARNING
Contact with the power supply voltage may cause an electric
shock, resulting in fatal or serious injury. Before checking
the voltage and fuse, make sure the power cord is
disconnected.
First check if the line voltage to be used matches the allowable line
voltage indicated on the label beside the power connector on the
spectrophotometer.
Fig. 1-2 Power Connector
Check if the fuse capacity is suitable for the line voltage to be used.
Table 1-1
Voltage
Fuse Capacity
Part No.
115 V
5 A (time lag type)
J821399
220 V/230 V/240 V
3.15 A (time lag type)
J821397
Fuse holder