Configuring the Contact Check Function
Using the contact check function
The contact check function improves the accuracy of potential measurement by detecting
incomplete probe contact. Ordinarily you should enable the function, although you can shorten
measurement times by disabling it. Change the setting as necessary.
•
Enable the contact check function.
Potential measurement
Analysis
Contact
check
Auto range
Measurement
Start of measurement
End of measurement
•
Disable the contact check function.
Potential measurement
Analysis
Auto range
Measurement
Start of measurement
End of measurement