37
Setting the Resistance Range
3.2 Setting the Resistance Range
Auto range
Select the optimal resistance range automatically (recommended).
Resistance range
When
[Auto range]
is disabled, allows you to select the resistance
range you wish to use. The measurement current (output current) is
displayed next to the resistance range.
1000 m
Ω
,
10
Ω
,
100
Ω
,
1000
Ω
,
10 k
Ω
,
100 k
Ω
,
1000 k
Ω
Choosing the range
Although
[Auto range]
selects the optimal resistance range automatically, measurement will take
more time than when the resistance range has been selected.
When measuring multiple electrode sheets with the same resistance value, perform the first
measurement using
[Auto range]
and then choose the resistance range that was used by the
[Auto
range]
function to perform the second and subsequent measurements in order to save time.
Change the setting as necessary.
•
Select
[Auto range
].
Potential measurement
Analysis
Contact
check
Auto range
Measurement
Start of measurement
End of measurement
•
Choose a resistance range.
Potential measurement
Analysis
Contact
check
Measurement
Start of measurement
End of measurement
3
Basic Measurement
Summary of Contents for RM2610
Page 2: ......
Page 12: ...8 Operating Precautions...
Page 22: ...18 USB License Key and Screens RM2612 Resistance Calculation Software...
Page 54: ...50 Checking Measurement Results...
Page 72: ...68 Setting Administrator Mode...
Page 110: ...106 Under what circumstances do contact errors occur...
Page 128: ...124 Dimensional Drawings 10 2 Dimensional Drawings RM2611 Electrode Resistance Meter Unit mm...
Page 130: ...126 Dimensional Drawings RM9005 Connection Cable Unit mm...
Page 134: ...130 Index...
Page 135: ......
Page 136: ......
Page 137: ......
Page 138: ......