108
Service life of replaceable parts
The performance of some parts used in the product may degrade during the course of extended
use. To ensure your ability to use the product into the indefinite future, Hioki recommends regular
replacement of these parts. Please contact your authorized Hioki distributor or reseller when you
need to do so.
Service life varies with the operating environment and frequency of use. Please note that
performance is not guaranteed for the duration of the recommended replacement cycle.
Part name
Recommended
replacement cycle
Remarks and conditions
RM2611 Electrode Resistance Meter
Electrolytic capacitors
About 10 years
The circuit board on which these
parts are mounted must be replaced.
LCD backlight (loss of half brightness)
About 50,000 hours
Backup battery
About 10 years
Replace when the RM2611 Electrode
Resistance Meter display shows
the following error code:
[ERR:096
Backup data error.]
.
Relay
About 50 million cycles
RM9004 Test Fixture
Probes
The recommended replacement cycle for probes depends on
the number of measurements made and the types of samples
measured. Please have the probes replaced once you start to
encounter frequent errors during SHORT and OPEN inspections
of the RM9004 Test Fixture.
Summary of Contents for RM2610
Page 2: ......
Page 12: ...8 Operating Precautions...
Page 22: ...18 USB License Key and Screens RM2612 Resistance Calculation Software...
Page 54: ...50 Checking Measurement Results...
Page 72: ...68 Setting Administrator Mode...
Page 110: ...106 Under what circumstances do contact errors occur...
Page 128: ...124 Dimensional Drawings 10 2 Dimensional Drawings RM2611 Electrode Resistance Meter Unit mm...
Page 130: ...126 Dimensional Drawings RM9005 Connection Cable Unit mm...
Page 134: ...130 Index...
Page 135: ......
Page 136: ......
Page 137: ......
Page 138: ......