79
RM2612 Resistance Calculation Software
Operating modes
Operating modes (initial value: potential measu analysis)
•
Potential measu analysis
•
Potential measurement
•
Analysis
Explanation of operation
•
Potential measu analysis: Instrument performs potential measurement and
then analysis.
•
Potential measurement: Instrument saves measured potential data to a file (one file
per measurement).
•
Analysis: Instrument loads a potential file saved during potential measurement,
analyzes the file, and saves the results to a file. Multiple files can be selected, and
analysis can be performed continuously.
Current error detection
The instrument detects current errors, aborts measurement, and recommends that
the user change the range.
Display: “ERROR”
Resistance over-range
detection
The instrument detects over-range events and recommends that the user change
the range.
Display: “ERROR”
Contact check function
Contact check ON/OFF
Resistance threshold: 1
Ω
to 1 M
Ω
(Initial values: ON, 20 k
Ω
[2.000E+04])
Input format: Decimal, exponential, up to 3 decimal places “0.000E+0”
•
Display: “CONTACT ERROR”
•
The instrument performs a simple measurement of the contact resistance between
the probes and the measurement target using two-terminal measurement and
concludes that a contact error has occurred if the resulting value is less than the
applicable threshold.
•
Contact error probe identification (estimation) function
The instrument combines measurements of the resistance between probes using
two terminals and identifies which probes triggered the contact error.
Measurement reliability
indicator function
Provides the following data for measured potential:
•
Consistency (legitimacy of the relative magnitude of center 5 points): PASS/FAIL
•
Variability per current channel: Current coefficient of variability [%]
•
Percentage of data rejected by the error data rejection function (error rate): [%]
(Measurement is stopped in the event of a FAIL result for consistency.)
Error data rejection
function
Rejection function ON/OFF (initial value: OFF)
Rejection of error data
•
Data with current errors is rejected.
•
Current channel data with a consistency result of FAIL is rejected.
•
Potential data from probes with contact errors is rejected from the calculation
process.
Counter function
Functionality for counting the number of measurements performed since the
application was launched (can be reset by user)
Counter values are output to analysis results files and potential files for analysis to
serve as a rough guide to the number of measurements performed by the RM9004
Test Fixture.
Comment entry function
Text entered into the comment field is linked to the measurement target and output
to a file.
1 comment per measurement (up to 30 characters)
10
9
8
7
6
5
4
3
2
1
索引
7
Specifications
Summary of Contents for RM2610
Page 2: ......
Page 12: ...8 Operating Precautions...
Page 22: ...18 USB License Key and Screens RM2612 Resistance Calculation Software...
Page 54: ...50 Checking Measurement Results...
Page 72: ...68 Setting Administrator Mode...
Page 110: ...106 Under what circumstances do contact errors occur...
Page 128: ...124 Dimensional Drawings 10 2 Dimensional Drawings RM2611 Electrode Resistance Meter Unit mm...
Page 130: ...126 Dimensional Drawings RM9005 Connection Cable Unit mm...
Page 134: ...130 Index...
Page 135: ......
Page 136: ......
Page 137: ......
Page 138: ......