36
Setting the Operating Mode
About operating modes
Potential measurement
Analysis
Meas.
Analysis
Meas. + Analysis
Start of measurement
End of measurement
You can choose the operating mode in order to streamline measurement work. When measuring a
small number of electrode sheets, it is recommended to set the mode to
[Meas.+Analysis]
so that
potential measurement and analysis are performed for each sheet.
When measuring multiple electrode sheets, efficiency suffers as the analysis step in
[Meas.+Analysis]
results in wait times. By choosing either
[Meas.]
or
[Analysis]
as the mode,
you can use the time it takes for analysis results to be generated more effectively by grouping
together analysis of multiple boards, boosting work efficiency.
This section describes the workflow when measuring multiple electrode sheets.
IMPORTANT
Each mode has its own
[File output]
setting. Before clicking
[Start]
, check the
[File output]
settings.
1
2, 7
6
3, 9
8
Potential measurement
1
Choose
[Meas.]
under
[Mode]
.
2
Click
[OK]
to accept the settings.
3
Place an electrode sheet in the RM9003 Press Unit and click
[Start]
to
perform potential measurement (p. 45).
4
A potential file for analysis use with a filename created by appending a time
stamp to the filename as set in
[Path (Analysis results)]
will be output. (One
file will be output for each measurement.)
5
Repeat the measurement step for each sheet (Step 3).
Analysis
6
Choose
[Analysis]
under
[Mode]
.
7
Click
[OK]
to accept the settings.
8
Select
[Analysis result]
under
[File output]
.
9
Click
[Start]
to display a file selection dialog box. Select multiple files output
during
[Meas.]
operation and click
[Open]
. (Analysis will be performed one
file at a time, and analysis results will be appended to the analysis results
file.)
Summary of Contents for RM2610
Page 2: ......
Page 12: ...8 Operating Precautions...
Page 22: ...18 USB License Key and Screens RM2612 Resistance Calculation Software...
Page 54: ...50 Checking Measurement Results...
Page 72: ...68 Setting Administrator Mode...
Page 110: ...106 Under what circumstances do contact errors occur...
Page 128: ...124 Dimensional Drawings 10 2 Dimensional Drawings RM2611 Electrode Resistance Meter Unit mm...
Page 130: ...126 Dimensional Drawings RM9005 Connection Cable Unit mm...
Page 134: ...130 Index...
Page 135: ......
Page 136: ......
Page 137: ......
Page 138: ......