103
What is measurement reliability?
8.9
What is measurement reliability?
Measurement reliability provides an indicator of whether potential measurement has been
performed correctly. It allows you to check the status of potential measurement. Measurement
reliability is indicated by three values.
Consistency [PASS/FAIL]
Consistency provides a judgment of whether the magnitude of the gradient in the potential
distribution accords with the theoretical result that would be expected. Potential increases near the
current source probe and decreases near the current drain probe, and consistency checks for the
relationship that would be expected in theory and expresses the result as a PASS or FAIL value.
Potential
Potential
Measurement
position
Measurement
position
Figure: Illustration of consistency (left: PASS consistency; right: FAIL consistency)
Error rate [%]
The error rate expresses the number of measurements with a incomplete contact as a percentage
of all measurements obtained during multipoint measurement of potential distribution. The error
rejection function rejects error data with an error rate that falls below a certain threshold and
performs analysis using only normal measurement data.
Coefficient of variability [%]
The coefficient of variability indicates the largest value of the coefficient of variability for equivalent
potential measurements. The coefficient of variability is defined as the result of dividing the standard
deviation by the average value. The instrument performs numerous equivalent measurements and
averages them, and the coefficient of variability for those measurements indicates the amount of
variation.
8
FA
Q
Summary of Contents for RM2610
Page 2: ......
Page 12: ...8 Operating Precautions...
Page 22: ...18 USB License Key and Screens RM2612 Resistance Calculation Software...
Page 54: ...50 Checking Measurement Results...
Page 72: ...68 Setting Administrator Mode...
Page 110: ...106 Under what circumstances do contact errors occur...
Page 128: ...124 Dimensional Drawings 10 2 Dimensional Drawings RM2611 Electrode Resistance Meter Unit mm...
Page 130: ...126 Dimensional Drawings RM9005 Connection Cable Unit mm...
Page 134: ...130 Index...
Page 135: ......
Page 136: ......
Page 137: ......
Page 138: ......