12
Operating Precautions
Operating Precautions
Observe the following precautionary information to ensure that the instrument can be used
safely and in a manner that allows it to perform as described in its specifications. Carefully
read the separate document entitled “Operating Precautions” before use. Use of the instrument
should conform not only to its specifications, but also to the specifications of all accessories,
options, and other equipment in use.
DANGER
Do not touch the section beyond the barrier during operation.
Failure to do so could cause the operator to experience an electric shock.
Do not measure any current in excess of the derating.
Doing so can cause overheating of the sensor, resulting in bodily injury, fire, or
damage to the instrument.
See “Frequency derating” (p. 65).
The maximum measurement current varies with the frequency, and the current
that can be measured continuously is limited. Operating the instrument at less
than this limitation is referred to as derating.
HIOKI CM4141C961-00