38
Current Measurement
Filter function
Filter off
Filter on
Measured value including noise
Measured value with reduced noise
Hold down
for 1 s.
Frequency (Hz)
Measured value (A)
Frequency characteristics when the filter function
is enabled
(100 A input)
Filter off
Filter on, passband 100 (Hz)
Disable the filter function when performing
measurement of power supply frequencies
in excess of 100 Hz, for example, on an
aircraft or ship.
Filter disabled
HIOKI CM4141C961-00
Summary of Contents for CM4141-50
Page 2: ...HIOKI CM4141C961 00 ...
Page 6: ...iv Contents HIOKI CM4141C961 00 ...
Page 24: ...18 Operating Precautions HIOKI CM4141C961 00 ...
Page 94: ...88 Accuracy table HIOKI CM4141C961 00 ...
Page 100: ...94 Cleaning HIOKI CM4141C961 00 ...
Page 104: ...Index 98 HIOKI CM4141C961 00 ...
Page 105: ...HIOKI CM4141C961 00 ...
Page 106: ...HIOKI CM4141C961 00 ...
Page 107: ...HIOKI CM4141C961 00 ...
Page 108: ...www hioki com HIOKI CM4141C961 00 ...