83
Accuracy table
Peak max, peak min
Range
Accuracy guarantee range
(resolution)
Accuracy guarantee
frequency range*
1
Measurement accuracy
6.000 V
0.00 V to ±12.00 V
(0.01 V)
10 Hz ≤ f < 45 Hz
±1.5% rdg ±0.07 V
DC, 45 Hz ≤ f ≤ 66 Hz
±1.0% rdg ±0.07 V
66 Hz < f < 1 kHz
±1.5% rdg ±0.07 V
60.00 V
±3.0 V to ±120.0 V
(0.1 V)
10 Hz ≤ f < 45 Hz
±1.5% rdg ±0.7 V
DC, 45 Hz ≤ f ≤ 66 Hz
±1.0% rdg ±0.7 V
66 Hz < f < 1 kHz
±1.5% rdg ±0.7 V
600.0 V
±30 V to ±1000 V*
2
(1 V)
10 Hz ≤ f < 45 Hz
±1.5% rdg ±7 V
DC, 45 Hz ≤ f ≤ 66 Hz
±1.0% rdg ±7 V
66 Hz < f < 1 kHz
±1.5% rdg ±7 V
1000 V
±50 V to ±1000 V*
3
(1 V)
10 Hz ≤ f < 45 Hz
±1.5% rdg ±7 V
DC, 45 Hz ≤ f ≤ 66 Hz
±1.0% rdg ±7 V
66 Hz < f < 1 kHz
±1.5% rdg ±7 V
*1: Accuracy within the frequency range of 10 Hz (inclusive) to 20 Hz (exclusive) is designed values.
*2: Values of up to ±1200 V are displayed, but accuracy is not defined for display values in excess of 1000 V
(values for reference purposes).
*3: Values of up to ±1500 V are displayed, but accuracy is not defined for display values in excess of 1000 V
(values for reference purposes).
HIOKI CM4141C961-00
Summary of Contents for CM4141-50
Page 2: ...HIOKI CM4141C961 00 ...
Page 6: ...iv Contents HIOKI CM4141C961 00 ...
Page 24: ...18 Operating Precautions HIOKI CM4141C961 00 ...
Page 94: ...88 Accuracy table HIOKI CM4141C961 00 ...
Page 100: ...94 Cleaning HIOKI CM4141C961 00 ...
Page 104: ...Index 98 HIOKI CM4141C961 00 ...
Page 105: ...HIOKI CM4141C961 00 ...
Page 106: ...HIOKI CM4141C961 00 ...
Page 107: ...HIOKI CM4141C961 00 ...
Page 108: ...www hioki com HIOKI CM4141C961 00 ...