86
Accuracy table
(12)
Diode
Range
Accuracy guarantee range
(resolution)
Short-circuit current
Measurement accuracy
1.800 V
0.000 V to 1.800 V*
1
(0.001 V)
200 µA ±20%
±0.7% rdg ±0.005 V
*1: A series of beeps indicates forward connection (0.15 V to 1.8 V). A continuous beep is emitted and the
LCD is backlighted in red if the voltage is less than 0.15 V.
(13)
Capacitance
Range (automatic ranging
threshold)
Accuracy guarantee range
(resolution)
Charging
current
Measurement
accuracy
1.000 µF
(more than 1100 counts)
0.000 µF to 1.100 µF
(0.001 µF)
10 nA ±20%
100 nA ±20%
1 µA ±20%
±1.9% rdg
±0.005 µF
10.00 µF
(more than 1100 counts,
less than 100 counts)
0.00 µF to 11.00 µF
(0.01 µF)
100 nA ±20%
1 µA ±20%
10 µA ±20%
±1.9% rdg
±0.05 µF
100.0 µF
(more than 1100 counts,
less than 100 counts)
0.0 µF to 110.0 µF
(0.1 µF)
1 µA ±20%
10 µA ±20%
100 µA ±20%
±1.9% rdg
±0.5 µF
1000 µF
(less than 100 counts)
0 µF to 1100 µF
(1 µF)
10 µA ±20%
100 µA ±20%
200 µA ±20%
±1.9% rdg
±5 µF
HIOKI CM4141C961-00