5
Introduction
Principle of leakage current measurement
The instrument is designed based on the
principle of electromagnetic induction. The
magnetic flux corresponding to the current
flowing through the conductor to be measured
is detected by a current transformer that
consists of a magnetic core and coil. The
current transformer generates the current
corresponding to the magnetic field (secondary
current). The detection resistor converts this
secondary current into the voltage to calculate
the value of the current flowing through the
conductor.
Magnetic core
Magnetic flux corresponding
to current being measured
Conductor under
measurement
Current to be
measured
Coil
Detection
resistor
Secondary
current