Inrush Current Measurement (AC INRUSH)
2.5 Inrush Current Measurement (AC INRUSH)
The AC inrush current can be measured.
Peak value of
AC inrush current
Screen example: 150
A
Inrush current occurrence
period
(Several tens to several
hundred of milliseconds)
AC current
RMS value (INRUSH)
Screen example: 60 A
For information on the trigger level, see the specifications. (p. 73)
An inrush current including DC components cannot be measured
accurately.
(Inrush current occurs.)
Auto-hold of measured value
INRUSH
will blink.
1
1
The inrush measurement range is set as follows in accordance with the
range when the current is measured.
Range when the current is measured INRUSH measurement range
AUTO
200.0 A range
6.000 mA, 60.00 mA
600.0 mA range
600.0 mA, 6.000 A, 60.00 A, 200.0 A
Range when the current is
measured
2
2
+
Hold down the key for
1 s or longer.
Switch over to AC
INRUSH mode.
Exit
+
How to exit AC INRUSH mode
Hold down the key
for 1 s or longer.