80
Input, Output, and Measurement Specifications
(2)
AC INRUSH (Inrush current)
1. AC INRUSH measured value
Range
Accuracy guarantee range
(Resolution)
Accuracy guarantee
frequency range
Measurement accuracy
600.0 mA
60.0 mA to 600.0 mA (0.1 mA)
40 Hz ≤ f ≤ 1 kHz
±5.0% rdg ±1.0 mA
6.000 A
±0.600 A to ±6.000 A (0.001 A)
40 Hz ≤ f ≤ 1 kHz
±5.0% rdg ±0.010 A
60.00 A
±1.00 A to ±60.00 A (0.01 A)
40 Hz ≤ f ≤ 1 kHz
±5.0% rdg ±0.10 A
200.0 A
±10.0 A to ±200.0 A (0.1 A)
40 Hz ≤ f ≤ 1 kHz
±5.0% rdg ±1.0 A
2. AC INRUSH PEAK value
Range
Accuracy guarantee range
(Resolution)
Accuracy guarantee
frequency range
Measurement accuracy
600.0 mA
60 mA to 1800 mA (1 mA)
40 Hz ≤ f ≤ 1 kHz
±6.0% rdg ±10 mA
6.000 A
±0.60 A to ±18.00 A (0.01 A)
40 Hz ≤ f ≤ 1 kHz
±6.0% rdg ±0.10 A
60.00 A
±1.0 A to ±180.0 A (0.1 A)
40 Hz ≤ f ≤ 1 kHz
±6.0% rdg ±1.5 A
200.0 A
±10 A to ±300 A (1 A)
40 Hz ≤ f ≤ 1 kHz
±6.0% rdg ±15 A
(3)
Frequency measurement
Range
(Auto-range threshold value)
Accuracy guarantee range
(Resolution)
Measurement accuracy
999.9 Hz (over 9999 counts)
15.0 Hz to 999.9 Hz (0.1 Hz)
±0.1% rdg ±0.1 Hz
2000 Hz (less than 900 counts)
900 Hz to 2000 Hz (1 Hz)
±0.1% rdg ±1 Hz
HIOKI CM4002A961-02
Summary of Contents for CM4002
Page 2: ...HIOKI CM4002A961 02...
Page 11: ...3 Introduction HIOKI CM4002A961 02...
Page 32: ...24 Display Panel HIOKI CM4002A961 02...
Page 40: ...32 Inspection Prior to Measurement HIOKI CM4002A961 02...
Page 100: ...Index 92 HIOKI CM4002A961 02...
Page 101: ...HIOKI CM4002A961 02...
Page 102: ...HIOKI CM4002A961 02...
Page 103: ...HIOKI CM4002A961 02...
Page 104: ...www hioki com HIOKI CM4002A961 02...