53
Output function (OUTPUT)
Setting the output function (RMS/WAVE)
Follow the instruction below to set the output function.
1
1
External power supply
2
2
(as required)
Recorder, etc.
CM4003
L9097
1
1
1
Connect the instrument and a device, such as a recorder,
using the L9097 Connection Cable (included).
2
Connect an external power supply as required. (p. 30)
3
If you enable the output function while using the auto-
range, the range is fixed at the presently set range.
4
Set the output function.
Hold down
the key for
1 s or longer.
1 s
RMS
WAVE
Off
RMS value
output
Waveform
output
Output
off
HIOKI CM4002A961-02