73
Input, Output, and Measurement Specifications
Frequency derating
characteristics
0
50
100
150
200
250
10
100
1 k
10 k
100 k
(A rms)
(Hz)
Crest factor
3 (other than 200.0 A range), 1.5 (200.0 A range)
Peak detection time width
2 ms or more (with filter disabled)
Frequency detection
input level
5% or more of f.s. in each range
Inrush trigger level
600.0 mA range
Not less than +60.0 mA (peak) or
not exceeding −60.0 mA (peak)
6.000 A range
Not less than +0.600 A (peak) or
not exceeding −0.600 A (peak)
60.00 A range
Not less than +2.00 A (peak) or
not exceeding −2.00 A (peak)
200.0 A range
Not less than +10.0 A (peak) or
not exceeding −10.0 A (peak)
HIOKI CM4002A961-02
Summary of Contents for CM4002
Page 2: ...HIOKI CM4002A961 02...
Page 11: ...3 Introduction HIOKI CM4002A961 02...
Page 32: ...24 Display Panel HIOKI CM4002A961 02...
Page 40: ...32 Inspection Prior to Measurement HIOKI CM4002A961 02...
Page 100: ...Index 92 HIOKI CM4002A961 02...
Page 101: ...HIOKI CM4002A961 02...
Page 102: ...HIOKI CM4002A961 02...
Page 103: ...HIOKI CM4002A961 02...
Page 104: ...www hioki com HIOKI CM4002A961 02...