6.1 Measurement Specifications
109
1
2
3
4
5
6
7
8
9
10
S
p
eci
ficat
io
ns
DIsplay Accuracy
REC Output Accuracy
(2 V/range)
MON Output Accuracy
(2 V/range)
Peak Display Accuracy
(Peak hold function)
L Range
H Range
When sine waves are continuously input after peak reset
±
(rdg.+dgt.)
DC
1 to 10 Hz
10 to 45 Hz 45 to 66 Hz 66 to 1 kHz
DC
± (0.3%+5)
--
--
--
--
AC
AC+DC
FAST
± (0.3%+7)
(AC+DC only)
± (0.3%+3)
± (0.3%+3)
NORMAL
± (0.3%+3)
SLOW
± (1.0%+3)
(AC+DC only)
±
(rdg.+mV)
DC
1 to 10 Hz
10 to 45 Hz 45 to 66 Hz 66 to 1 kHz
AC+DC
FAST
± (0.8%+10)
(AC+DC only)
--
--
± (0.8%+5)
± (0.8%+5)
NORMAL
± (0.8%+5)
SLOW
± (1.3%+5)
(AC+DC only)
±
(rdg.+mV)
DC
1 to 10 Hz
10 to 45 Hz 45 to 66 Hz 66 to 1 kHz
DC
AC+DC
± (0.3%+5)
± (0.3%+5)
± (0.3%+5) ± (0.3%+5)
± (0.3%+5)
AC
--
--
±
(rdg.+dgt.)
DC
1 to 10 Hz
10 to 45 Hz
45 to 66 Hz
66 to 1 kHz
DC
AC+DC
± (0.5%+2)
± (0.5%+2)
± (0.5%+2)
± (0.5%+2)
± (0.5%+2)
AC
--
--
±
(rdg.+dgt.)
DC
1 to 10 Hz
10 to 45 Hz
45 to 66 Hz
66 to 1 kHz
DC
AC+DC
± (0.5%+5)
± (0.5%+5)
± (0.5%+5)
± (0.5%+5)
± (0.5%+5)
AC
--
--