6.3 CLAMP ON AC/DC SENSOR Specifications
6.3.3 Model 9693 CLAMP ON AC/DC SENSOR
Rated current
AC/DC 2000 A
Output voltage
AC/DC 2 V/ 2000 A
Output resistance
100
Ω
or less
Maximum permissible
input
2000 Arms (Derating according to frequency)
Maximum peak current
value
2840 A
Continuous input
❖
See "Continuous Input" (page 121)
Peak input
❖
Frequency band
DC to 15 kHz (-3dB)
Effect of conductor
position
Within ± 0.7%
Effect of external
electromagnetic field
2.0 A equivalent or less
(in an external electromagnetic field of 400 A/m)
Temperature
characteristics
0 to 40
°
C range: 0.1 x accuracy specifications/
°
C
32 to 104°F range: 0.18 x accuracy specifications/°F
Location for use
Indoor, altitude up to 2000 m (6566 feet)
Operating
temperature and
humidity range
0 to 40°C (32 to 104°F), 80% RH or less (no condensation)
Storage temperature
and humidity
-10 to 50°C (14 to 122°F), 80% RH or less (no condensation)
Operating temperature
and humidity for
guaranteed accuracy
Guaranteed
accuracy period
23 ±
5°C (73 ± 9°F), 80% RH or less (no condensation)
For one year (Opening and closing of the sensor: 10,000 times)
Maximum permitted
circuit voltage
600 V
Dielectric strength
5.55 kV AC for 1 minute (sensor - case, sensor - circuit)
Power supply
± 5 V ± 0.25 V (supply from the outside)
Maximum rated power
50 mVA
Measurable conductor
diameter
55 mm (2.17") or less
Dimensions
Approx. 62W x 196H x 35D mm
(Approx. 2.44"W x 7.72"H x 1.38"D)
Mass
Approx. 500 g (17.6 oz.)
Cable length
Approx. 2 m (6.5 feet)