Page 64
171051(12) - Hillrom™ 900 Electric Bed Instructions for Use
Oxygen Cylinder Holder (AC959A-AD101A-AD102A)
Safe working load: 15 kg
(1)
The oxygen cylinder holder is designed to accept an oxygen cylinder and must
only be fitted on the patient helper supports at the head end of the bed outside the
sleep surface. It can be rotated through 80°. Each type of holder corresponds to a
cylinder model and must never be used with a different cylinder. See below.
AC959A for
cylinder model
B5 (Ø140)
AD101A for
cylinder model D
(Ø100)
AD102A for
cylinder model E
(Ø100)
The following recommendations are designed to prevent any possible incidents so that
this accessory can be used in optimum safety conditions for both the patient and
nursing staff.
• Check that the cylinder is correctly
positioned at the base of the cylinder
holder.
• Never use a different oxygen cylinder
model from the model that is specified
above (danger of dropping the cylinder
or interfering with various operations
could occur).
• Prevent any impact when moving a
bed equipped with a cylinder holder
(especially doorways).
• If the cylinder holder does not allow
the bed to go through a doorway,
position the holder in front of the bed,
otherwise place it and the cylinder on the mattress (remember to put the
holder in its normal position after moving the bed).
1. The safe working load specifications allow for a substantial safety margin.