171051(12) - Hillrom™ 900 Electric Bed Instructions for Use
Page 9
Int
rodu
ct
ion
Spe
c
if
ic
at
ions
Never clean or service the bed without unplugging it from the mains power supply
and disconnecting the battery.
The battery backup must never be left in direct contact with fire, placed in liquid, or
discarded in a refuse bin. In the event of the battery being damaged, see “De-
commissioning” page 80.
This label indicates that the bed
must never be used with an
oxygen tent or in explosive atmospheres
(presence of
inflammable gases or vapors)
.
Use only nasal tubes and oxygen
masks. For reasons of safety, masks and tubes should always be
kept at a higher level than the sleep surface.
Always lock out the HiLow function before any cleaning or
maintenance operations.
If the bed is equipped with a battery, and the bed is stored for long periods of time, the
battery must be charged every 3 months. Failure to do so could result in damage to the
battery.
If a continuous alarm sounds when activating a movement, the battery needs to
be recharged.
General precautions for the place of use
It is advisable not to use the bed under the following conditions:
- in hospital wards other than the intended ward (see “Intended Use” page
- climatic conditions outside the corresponding ranges recommended by
Hill-Rom,
- in hyperbaric chambers,
- in explosive atmospheres,
- in the presence of flammable gases or vapors,
- with oxygen tent type respiration devices or devices that extend below
the sleep surface,
- outdoors or to transport a patient in a vehicle,
- moving the bed over soft ground or inappropriate surfaces,
- moving the bed along slopes of over 10° (with or without a patient).
Service temperature
10° and +40°
Service humidity
30% - 85%
Working atmospheric pressure
700 hPa to 1,060 hPa
Climatic restrictions