Page 46
171051(12) - Hillrom™ 900 Electric Bed Instructions for Use
2.
Check that neither the mattress nor any accessories are touching the fixed
parts of the bed (and in particular the head section) and that no traction is
applied to the parts installed above and below the sleep surface (e.g.,
power cable and air mattress pipes).
Correct position
Incorrect position
Headboard
3.
Install the mattress (only mattresses recommended by Hill-Rom (See table
on page 26), cushions, sheets and blankets, and all other accessories that
must remain on the bed.
The weight of these additional articles must not exceed 65 kg or 45 kg,
depending on the destination of the product and the maximum patient weight
(see “Technical specifications” page 11). No more than 39 kg must be added at
a time.
The system is initialized without the patient on the bed.
1.
Press the initialization button and hold until the bed reaches
the initialization position (sleep surface in the raised
position and horizontal). A beep indicates that initialization
can begin. Release the button
2.
The orange “do not touch” light flashes during initialization.
3.
A beep sounds and the light turns green to indicate that the
initialization is complete, then it goes out after a few
seconds.
NOTE:
If the initialization fails, a beep sounds three times and the light
turns orange.
NOTE:
If a load is added to the bed and the orange light comes on, it is
necessary to proceed with an initialization.