Classification and Standards
The Advanta™ 2 Bed is designed and manufactured according to the following equipment classifications and
standards:
Electromagnetic Emissions Guidance
Guidance and Manufacturer's Declaration—Electromagnetic Emissions
The Advanta™ 2 Bed model P1190 is intended for use in the electromagnetic environment specified below. The
customer or the user of the model P1190 should make sure it is used in such an environment.
Emissions Test
Compliance
Electromagnetic Environment—Guidance
RF Emissions
CISPR 11
Group 1
The model P1190 uses RF energy only for its internal func-
tions. Therefore, its RF emissions are low and are not likely to
cause any interference in nearby electronic equipment.
RF Emissions
CISPR 11
Class A
The model P1190 is suitable for use in all establishments other
than domestic and those directly connected to the public low-
voltage power supply network that supplies buildings used for
domestic purposes.
Harmonic Emissions
IEC 61000-3-2
Not
Applicable
Voltage Fluctuations/
Flicker Emissions
IEC 61000-3-3
Not
Applicable
Technical and Quality Assurance Standards
UL 60601-1
CSA® C22.2 No. 601.1
IEC 60601-2-38
IEC 60601-1
IEC 60601-1-2
EN ISO 9001 and EN 13485
Equipment Classification per IEC 60601-1
Class
I
equipment, internally powered equipment
Classification according to EU Directive 93/42/EEC
Class
I
Degree of Protection Against Electric Shock
Type B
Degree of Protection Against Ingress of Water
Protection against spraying or splashing water- IPX4
Degree of Protection Against the Presence of
Flammable Anaesthetic Mixtures
Not for use with flammable anaesthetics.
Mode of Operation
Continuous operation with intermittent loading,
3 minutes ON/27 minutes OFF
Sound level (except alarms) (measured 1 meter from
patient’s ear)
< 60 dBA for normal operation, does not include the
IntelliDrive® Transport System or transients
< 70 dBA maximum (IntelliDrive® Transport System active)
< 78 dBA transients (brake/steer activation, siderail latch and
unlatch)