26
Advanta™ 2 Bed User Manual (157722 REV 2)
2. Press, and hold, these controls at the same time for five seconds:
Knee Up
,
Knee Down
,
Bed Up
, and
Bed Down
. The Alerts will flash green for three seconds to let you know the configuration is set.
To make sure the Alerts operate as configured, do as follows:
1. Put the siderails in the applicable configuration.
2. Make sure the Alerts are green.
3. Lower one of the configured siderails, and make sure the Alerts flash yellow. Then, raise the same
siderail, and make sure the Alerts are green.
4. Repeat step 3 for each of the configured siderails.
NOTE:
If more than the configured siderails are
up
, the system will not alert. It only alerts when the configured
siderails are
down
.
Optional Patient Controls
The optional patient controls are located on the patient pendant.
Nurse Call Control
Nurse
Answer
Nurse
Call
The Nurse Call control is located on the inside of the head siderails and on the
patient pendant.
To Activate
•
Press any
Nurse Call
control.
•
When the nurses station acknowledges the nurse call, the
Nurse Call
LED
comes on.
•
When the nurses station’s communication line is open, the
Nurse Answer
LED
comes on.
•
Speak into the speaker/microphone located on the inside of the head siderails.
Room Light Control
The room light control allows the patient to turn the room light off and on.
Reading Light Control
The reading light control allows the patient to turn the reading light off and on.
Volume Control
The volume control allows the patient to adjust the volume of the television or radio in the
room.