Service Manual LikoGuard™
overhead lift
• 3EN190401 Rev. 7
8
www.hillrom.com
2.3 Error codes - Service Advice
HandControl LG with display
Error
code
Meaning:
Recovery:
Service advice:
0310
Motor current inrush error
5 minute time out before the system
is operational again
If error repeats; Contact Hill-Rom
0311
Current difference greater than
5Amps between the motors
5 minute time out before the system
is operational again
If error repeats; Contact Hill-Rom
0314
Potentiometer has failed
5 minute time out before the system
is operational again
Check :
1. Potentiometer connects to the gear
wheel
2. Electrical connection to the
potentiometer
3. Electrical connection to the board
4. Continuity of the cable from
potentiometer to board
0318
Invalid current read when lift
not moving
5 minute time out before the system
is operational again
If error repeats; Contact Hill-Rom
0319
Overtemperature occurred
System does not allow operation if
the temperature is too high. When
the temperature cools down the
system can operate again
If error repeats; Contact Hill-Rom
0320
Error occurred during motor
calibration
System can function but using
default motor calibration values
Contact Hill-Rom
0321
Overload occurred detected by
the software, should only occur
if lifting over the SWL,
(Software trigger)
Lift should allow downward move
-
ment, upward movement available
after 2 seconds
If this occurs at loads below the SWL;
contact Hill Rom.
If this occurs above the SWL, the lift is
being used out of specification.
0322 Overload occurred detected by
the software, should only occur
if lifting over the SWL,
(Hardware trigger)
Lift should allow downward move
-
ment, upward movement available
after 2 seconds
If this occurs at loads below the SWL;
contact Hill Rom.
If this occurs above the SWL, the lift is
being used out of specification.
0430
External memory failure -
generates a service request*
Should recover automatically
Check parameters and reset service**
0431
External memory failure -
generates a service request*
Should recover automatically
Check parameters and reset service**
0605
Bus communication error
If error repeats; Contact Hill-Rom
0606
Bus communication error
If error repeats; Contact Hill-Rom
0607
Bus communication error
If error repeats; Contact Hill-Rom
0609
Bus communication error
If error repeats; Contact Hill-Rom
0630
External memory failure -
generates a service request*
Should recover automatically
Check parameters and reset service **
0631
External memory failure -
generates a service request*
Should recover automatically
Check parameters and reset service**
060b
Bus communication error
If error repeats; Contact Hill-Rom
060c
Bus communication error
If error repeats; Contact Hill-Rom
090d
RTC (real time clock) is
desynchronised, service flag
should be set
Replace the Main Board battery (button),
reset service **
*
For the HandControl without display service request is indicated by the wrench icon LED.
For the Hand Control with display service request is indicated by the display icon (top right).
** To reset service, see 4.7