www.hillrom.com
+
Symbols and Information (pop-up message)
A pop-up message is displayed on screen for 5 seconds or until any button on the HandControl is pressed.
Service required
No lifting allowed!
Low battery level
Overload
operating time
Low battery!
The battery should
be charged at once.
Broken contact!
HandControl has lost
signal from the lift.
No lifting allowed!
A serious operational fault
has developed. Lifting is not
permitted.
Emergency stop activated!
Press in the emergency
stop button on the
underside of the lift to
reset.
T
o Menus!
To display menus
Usage or Service;
Hold in the button
on the
HandControl for 2
seconds.
Service required
The symbol is continuously
illuminated
yellow.
-The lift may be used as
normal until Service.
Overload
(maximum load)!
The symbol is illuminated
yellow, an alarm sounds
and the lift stops.
-Check load, and that
the lift is not affected by
surrounding objects.
When the symbol goes
out and the alarm goes
off the lift can be used
again.
Overload
(operating time)!
The symbol flashes yellow.
- The lift’s operating time has
been exceeded. The lift may be
used to lower the patient to a
secure surface. Allow the lift
to rest for around 2 hours or
until the message is no longer
displayed when the button is
pressed.
When the symbol goes out the
lift can be used again.
Overload
press
Open menus
2 seconds
3.3 Pop-Up messages
•
After 1 minute of unuse (all buttons), back to password
and continue the navigation through different screens.
•
After 5 minutes of unuse (all buttons), back to screensaver.
•
If up or down buttons are used during the screen navigation,